■ APSA-380 Specifications
| Component | Sulfur dioxide (SO2) |
|---|---|
| Measuring principle | Ultraviolet Fluorescence |
| Measurement range | 0-0.05/0.1/0.2/0.5/1.0/5.0/10/20 ppm (up to 8 ranges selectable, auto-range)*1 |
| Lower Detection Limit (2σ) | < 0.3 ppb*2*3 |
| Measurement unit | ppb, µg/m3, ppm, mg/m3 |
| Sample gas flow rate | Approx. 0.6 L/min |
| Repeatability | ±1.0% of full scale |
| Linearity | ±1.0% of full scale |
| Zero drift | < 0.5 ppb (24 h) |
| Span drift | < 0.5% of full scale (24 h) |
| Response time (T95) | < 120 s from the inlet*3 |
| Display | 7-inch color LCD with touch panel |
| Communication | Ethernet × 2 (Modbus® TCP), RS-232C × 1 (Modbus® RTU) , USB flash drive × 1 |
| Analog output (option) | Maximum : 3 channels (insulated) DC4-20 mA, DC 0-0.1 V, DC 0-1 V, DC 0-5 V or DC 0-10 V |
| Digital output (option) | Relay contact output for range 3 channels, relay contact output for other 6 channels |
| Digital input (option) | Maximum : 4 channels, non-isolated input |
| Installation environment | Operation temperature : 0-40℃ (32℉ - 104℉), relative humidity : 85% or less |
| Temperature and pressure compensation | With |
| Sampling pump and filter | Internal |
| Power supply | AC 100-240V ±10% (max. voltage: AC 250 V), 50/60 Hz, consumption 50 W |
| Dimensions, weight | 430 (W) x 568 (D) x 221 (H) mm, Approx. 18 kg |
| Compliance | CE, UKCA, KC, FCC, China RoHS |
*1 : 2nd point calibration is required in case of using higher than 1 ppm range *2 : For ranges less than 1 ppm *3 : Digital filter
・ If there is other measurement range requirement than shown above, please consult HORIBA
・ Modbus is a trademark of Schneider Electric USA Inc.
■ Dimensional Outline (Unit: mm)

■ Rear Panel Configuration

1. Exhaust 6. I/O terminal block
2. Calibration gas inlet (option) 7. USB flash drive connector
3. Sample gas inlet 8. Ethernet connector
4. Guide label 9. RS-232C connector
5. AC power connector
■ Flow Sheet

■ Options
▪ I/O terminal block (Analog input/output, digital input/output)
▪ Pressurized calibration unit ▪ Flow sensor ▪ USB flash drive
▪ Mounting parts (rubber feet, brackets and slide rails, brackets for slide rails)

Ultimate Performance
User-Friendly Interface & Functionality
Reduced Operational Expense & Maximum Uptime
Connected & Intelligent
Module Design for Various Needs & Customizations
Environment-Friendly Design